Jan 5, 2016Physical SciencesAAS227 20160 likes1 view Auto-indexed

Verification of Absolute Calibration of Quantum Efficiency for LSST CCDs

Keywords:LSSTAstronomyCCD CamerasInstrumentationQuantum EfficiencyLSST CCDsCharge-Coupled Devicesdark energysupernovae
Poster thumbnail

Description

We describe a system to measure the Quantum Efficiency in the wavelength range of 300nm to 1100nm of 40x40 mm n-channel CCD sensors for the construction of the 3.2 gigapixel LSST focal plane. The technique uses a series of instruments to create a very uniform flux of photons of controllable intensity in the wavelength range of interest across the face of the sensor. This allows the absolute Quantum Efficiency to be measured with an accuracy in the 1% range. This system will be part of a production facility at Brookhaven National Lab for the basic components of the LSST camera.

Conference Information

American Astronomical Society, AAS Meeting #227

Acronym:AAS227
Year:2016
Location:Kissimmee, United States
Dates:Jan 4 - Jan 8, 2016

Identifiers

Authors

Rebecca Coles

Wayne State University / Brookhaven National Laboratory

James Chiang

SLAC National Accelerator Laboratory

David Cinabro

Wayne State University

Woodrow Gilbertson

Purdue University

Justine Haupt

Brookhaven National Laboratory

Ivan Kotov

Brookhaven National Laboratory

Homer Neal

SLAC National Accelerator Laboratory

Andrei Nomerotski

Brookhaven National Laboratory

Paul O'Connor

Brookhaven National Laboratory

Christopher Stubbs

Harvard University

Peter Takacs

Brookhaven National Laboratory

Additional Information

LanguageEnglish
Formatapplication/pdf
Version1
License
Creative Commons Attribution 4.0 International
Indexed fromZenodo