Description
MicroCHANDLER is a prototype detector for surface-level reactor antineutrino detection made with plastic scintillator cubes (EJ-260). We measured the proton quenching effect in the scintillator with neutron beams of 5-27 MeV at the Triangle Universities Nuclear Laboratory (TUNL). We extracted the Birks' constants from our data. This result improves the energy non-linearity modeling of plastic scintillator detectors in an unprecedented energy range. In addition, the updated energy response function for recoil protons will lead to an improved modeling of fast neutrons, which is needed to optimize the shielding design.
Conference Information
Neutrino 2020
Identifiers
Authors
Keegan Walkup
Alireza Haghighat
Tulasi Subedi
Samuel Hedges
Jonathan Link
Jay Runge
Valerio Mascolino
Connor Awe
Shengchao Li
Virginia Tech
Jacob Steenis
Phillip Barbeau
Tyler Johnson