Jun 24, 2020Physical Sciences0 likes Auto-indexed

Improvement of Energy Resolution in KamLAND-Zen by Implementing Signal Amplifier for Low-gain PMTs

Keywords:Neutrinoless Double Beta DecayKamLAND-Zensignal amplifier
Poster thumbnail

Description

KamLAND-Zen is a neutrinoless double-beta(0ν2β) decay search experiment using $^{136}$Xe installed in the KamLAND detector . KamLAND mainly consists of 1 kt liquid scintillator and 1,879 photomultiplier tubes(PMTs). One of the dominant background sources in the 0ν2β decay energy region around the Q-value of 2.459 MeV is 2ν2β decay due to the finite energy resolution. Since the only way to separate 0ν2β decay from 2ν2β decay is energy information, the energy resolution is important in the high sensitive search. However, the number of unused PMTs is increasing due to the decreasing gains. It makes the energy resolutoin of KamLAND worse. To recover the energy resolution, the signal amplifiers for such "low-gain PMTs" were developed and installed. I will report the recovery of the PMT performance and the improvement of the detector energy resolution by the signal amplifiers.

Conference Information

Neutrino 2020

Year:2020

Identifiers

Authors

Nanami Kawada

Tohoku University

Additional Information

LanguageEnglish
Formatapplication/pdf
Version1
License
Creative Commons Attribution 4.0 International
Indexed fromZenodo